SPECIFICATIONS |
|
Specifications apply after 30 minute
warm-up, ambient 5°C to 40°C |
|
FREQUENCY |
|
Frequency Range: |
150kHz
to 2000MHz |
|
Setting
Resolution: |
10Hz by
direct keyboard entry, or in
user-set increments of 10Hz to
999·999MHz by rotary control or
increment–decrement keys. |
|
Display
Resolution: |
10Hz |
|
Phase
Noise: |
-116dBc/Hz at 25kHz offset, 500MHz
carrier. |
|
Residual
FM (FM Off): |
Equivalent peak deviation for 300Hz
to 3·4kHz B/W: 10Hz at 500MHz
carrier |
|
REFERENCE FREQUENCY |
|
Options: |
Internal
or External (via rear panel BNC) |
|
Internal
Accuracy: |
+/- 1
ppm over temperature range 1 5°C to
30°C; +/- 2 ppm over 5°C to 40°C. |
|
Internal
Stability: |
<
+/-1ppm/year ageing. |
|
Internal
Ref. Out: |
10MHz
from 50 Ohms, amplitude 2V pk-pk
into 50 Ohms. |
|
External
Ref In: |
10MHz
into 50 Ohms, amplitude 2V pk-pk to
5V pk-pk. |
|
OUTPUT LEVEL |
|
Output
Level Range: |
–127dBm
to +7dBm (0·1µV to 500mV into 50W).
–127dBm to +1dBm in AM mode. |
|
Setting
Resolution: |
0·1dB
(or 0·01µV to 1mV) by direct
keyboard entry, or in user-set
increments of 0·1dB to 100dB (or
0·01µV to 100mV) by rotary control
or increment–decrement keys. |
|
Accuracy: |
Better
than ± 2dBm. |
|
Harmonics: |
<-25dBc
at +7dBm. |
|
Sub-Harmonics: |
<-25dBc
at +7dBm. |
|
Non-Harmonic Spurii: |
<–60dBc
at >62.5MHz, <62.5MHz.<–50dBc at <62.5MHz. |
|
Carrier
Leakage: |
<0.5uV
generated into a 50W load by a 2
turn 25mm loop, at 25mm from the
generator with output set to <-10dBm
into a 50W sealed load. |
|
Output
Impedance: |
50W |
|
Output
Connector: |
TYPE N |
|
Reverse
Protection: |
50V DC,
up to 25W from 50 Ohm source, LED
indication |
|
Output
Switch: |
RF OUT
on-off switch with LED for ON
status. |
|
MODULATION SOURCE |
|
Type: |
Internal
from built-in sine wave generator,
or external from front panel BNC. |
|
Internal: |
400Hz or
1kHz sine, signal also available as
an output. |
|
External: |
Calibrated for 1V rms sine, input
impedance 600 Ohms. |
|
FREQUENCY MODULATION |
|
Max.
Peak Deviation: |
See
Table below. |
|
Setting
Resolution: |
0·5kHz
by direct keyboard entry, rotary
control or increment–decrement keys. |
|
Deviation Accuracy: |
|
|
External
Modulation |
100Hz -
300kHz (± 2dB relative to 1kHz). |
|
Distortion: |
< 2% at
1kHz modulation, max. deviation
(300-3.4kHz bandwidth). |
|
PHASE
MODULATION |
|
Max.
Peak Deviation: |
See
Table below. |
|
Setting
Resolution: |
0.05
rads for <10.0rads deviation, 0.1
rads for >10.0rads deviation. |
|
Deviation Accuracy: |
|
|
External
Modulation |
100Hz -
10kHz (±2dB relative to 1kHz). |
|
Distortion: |
< 2% at
1kHz modulation, max. deviation
(300-3.4kHz bandwidth). |
|
AMPLITUDE MODULATION |
|
Max Mod.
Depth: |
100%. |
|
Setting
Resolution: |
0.5%. |
|
Deviation Accuracy: |
<± (5%
setting +1%) for 1kHz Internal or
1kHz / 1Vrms External Modulation. |
|
External
Modulation |
50Hz -
200kHz (±1dB relative to 1kHz). |
|
Distortion: |
150kHz
to 1GHz - <3% at 30%, <5% at 70%1GHz
to 2GHz - <5% at 30%,
<10% at 70% at 1kHz modulation, max.
deviation (300-3.4kHz bandwidth). |
|
INTERFACES |
|
Full
remote control facilities are
available through both RS232 and
GPIB interfaces |
|
RS232: |
Variable
Baud rate, 19200 Baud maximum, 9-pin
D-connector. Compatible with Thurlby
Thandar ARC (Addressable RS232
Chain). |
|
GPIB
(IEEE-488): |
Conforming with IEEE488·1 and
IEEE488·2. |
|
GENERAL |
|
Display: |
20
character x 4 row backlit
alphanumeric LCD |
|
Data
Entry: |
Keyboard
selection of frequency, amplitude,
etc.; value entry by numeric keys or
by rotary control. |
|
Stored
Settings: |
Up to 9
complete instrument set-ups may be
stored in battery-backed memory. |
|
Size: |
3U
(130mm) height; half-rack (212mm)
width; 330mm long. |
|
Weight: |
4·6 kg.
(10 lb) |
|
Power: |
230V,
115V or 100V nominal 50/60Hz,
adjustable internally; operating
range ±14% of nominal; 30VA max.
Installation Category II. |
|
Options: |
19 inch
rack mounting kit. |
|
ENVIRONMENTAL & SAFETY |
|
Operating Range: |
+5°C to
40°C, 20-80% RH. |
|
Storage
Range: |
–20°C to
+ 60°C. |
|
Environmental: |
Indoor
use < 2000m, Pollution Degree 1. |
|
Safety: |
Complies
with EN61010-1. |
|
EMC: |
Complies
with EN61326 |